Instructions for use:
MANUAL CLEANING INSTRUCTUIONS
1. Ensure to remove all instruments from the Dentistry Lock Grab Pad prior to cleaning the pad.
2. Rinse the Dentistry Lock Grab Pad under running, cold utility water to remove excess contamination.
3. Prepare a detergent bath, such as Dawn® dish soap, at a minimum concentration of 1 oz/10 gallon (~3 mL/gallon) using warm utility water.
4. Fully immerse the Dentistry Lock Grab Pad in the prepared detergent bath and allow the pad to soak for a minimum of 2 minutes. While soaking, wipe down the device with lint-free cloths soaked in dish soap solution to remove visible soil.
5. Remove the Dentistry Lock Grab Pad from the prepared dish soap bath and rinse it under running critical water to remove dish soap residuals.
6. Pat the Dentistry Lock Grab Pad dry using lint-free cloths.
7. Visibly inspect the Dentistry Lock Grab Pad for contamination. Repeat steps 2 through 7 if contamination is observed.
STERILIZATION PARAMETERS
Sterilizer Type: Pre-Vacuum
Temperature: 132°C
Pulses: 4
Fully Cycle Exposure Time: 4 minutes
Dry Time: 60 minutes
Test Article Configuration: Wrap the Dentistry Lock® Pad in two layers of Halyard Health H200 (510(k) K0882554), or equivalent, polypropylene wrap using simultaneous envelope folding techniques.